A design study of the energy selection system for carbon-ion therapy

Garam Hahn, Dong Hyun An, Bong Hwan Hong, Geun Beom Kim, Heejoong Yim, Hong Seok Chang, In Su Jung, Kun Uk Kang, Sang Hoon Nam, Inkyu Park

Research output: Contribution to journalArticlepeer-review

3 Scopus citations

Abstract

KHIMA, a research project to construct a carbon radio-therapy facility in Korea, has been developing a superconducting cyclotron named KIRAMS-430 as a carbon(12C6+) particle accelerator. Due to the fixed beam energy of the cyclotron, an energy selection system (ESS) is required for treatment of tumors located at various depths in the human body. In the present paper, two design stages of the ESS are discussed. First, the beam tracks behind the degrader block and the statistical twiss parameters for the entire energy range were calculated by using the GEANT4 simulation toolkit. Analysis of the beam transmission and the contamination ratios were performed. In the second stage, the beam optics was designed to support the same phase profile at the end regardless of the variations in all of input twiss parameters and the emittance.

Original languageEnglish
Pages (from-to)513-517
Number of pages5
JournalJournal of the Korean Physical Society
Volume66
Issue number3
DOIs
StatePublished - 2015

Keywords

  • Cyclotron
  • ESS
  • KHIMA
  • KIRAMS-430
  • Medical

Fingerprint

Dive into the research topics of 'A design study of the energy selection system for carbon-ion therapy'. Together they form a unique fingerprint.