Abstract
Objectives: To systematically explore the lowest reasonably achievable radiation dose for appendiceal CT using an iterative reconstruction (IR) in young adults. Methods: We prospectively included 30 patients who underwent 2.0-mSv CT for suspected appendicitis. From the helical projection data, 1.5-, 1.0- and 0.5-mSv CTs were generated using a low-dose simulation tool and the knowledge-based IR. We performed step-wise non-inferiority tests sequentially comparing 2.0-mSv CT with each of 1.5-, 1.0- and 0.5-mSv CT, with a predetermined non-inferiority margin of 0.06. The primary end point was the pooled area under the receiver-operating-characteristic curve (AUC) for three abdominal and three non-abdominal radiologists. Results: For the abdominal radiologists, the non-inferiorities of 1.5-, 1.0- and 0.5-mSv CT to 2.0-mSv CT were sequentially accepted [pooled AUC difference: 2.0 vs. 0.5 mSv, 0.017 (95% CI: -0.016, 0.050)]. For the non-abdominal radiologists, the non-inferiorities of 1.5- and 1.0-mSv CT were accepted; however, the non-inferiority of 0.5-mSv CT could not be proved [pooled AUC difference: 2.0 vs. 1.0 mSv, -0.017 (-0.070, 0.035) and 2.0 vs. 0.5 mSv, 0.045 (-0.071, 0.161)]. Conclusion: The 1.0-mSv appendiceal CT was non-inferior to 2.0-mSv CT in terms of diagnostic performance for both abdominal and non-abdominal radiologists; 0.5-mSv appendiceal CT was non-inferior only for abdominal radiologists. Key points: • For both abdominal and non-abdominal radiologists, 1.0-mSv appendiceal CT could be feasible. • The 0.5-mSv CT was non-inferior to 2.0-mSv CT only for expert abdominal radiologists. • Reader experience is an important factor affecting diagnostic impairment by low-dose CT.
Original language | English |
---|---|
Pages (from-to) | 1826-1834 |
Number of pages | 9 |
Journal | European Radiology |
Volume | 28 |
Issue number | 5 |
DOIs | |
State | Published - 1 May 2018 |
Keywords
- Appendicitis
- Prospective studies
- ROC curve
- Sensitivity and specificity
- Tomography, X-ray computed